Explore the Interface

logo Additel

รุ่น 780


Description Additel 780

Pressure Controller
  • Pressure ranges from vacuum to 3,000 psi (200 bar)
  • Precision accuracy models to 0.01% of reading
  • Standard model accuracy of 0.02% of full scale
  • HART Communication and Profibus PA
  • Control stability of 0.003%FS
  • Easy-to-use icon based user interface
  • ISO 17025-accredited calibration w/data included


  • For years, we’ve provided the most durable, accurate, quality pressure calibration products for field applications. The Additel 780 controller incorporates the same durability, accuracy, and quality into a new bench top controller packed with features and functionality that is remarkably easy to use. The Additel 780 offers two base ranges: to 1,000 psi (70 bar) and to 3,000 psi (200 bar). The base range establishes the maximum controlling range of the controller. Each configuration includes a control sensor in which you can preselect the sensor range best suited your application. External and internal sensors can be used which allows for expanded range and accuracy capability in the future.

    Modular Design
    Each unit comes with one Intelligent Pressure Module configured to the many range offerings provided. Standard accuracy sensors (ADT160A-CPXXX) are silicon pressure sensors with a 1 year accuracy of 0.02% FS. The precision quartz based sensors improve the 1 year accuracy specification to 0.005% of reading + 0.005% FS. Each sensor has been specially aged, tested and screened before assembly. After assembly each sensor is temperature compensated over the range of -10°C to 50°C. The Additel 780 allows for one internal pressure sensor and one external pressure sensor. The modular design of this unit provides for interchangeability of both the internal and external sensors with other Intelligent Pressure Modules.In addition to the Intelligent Pressure Modules, the Additel 780 has a built in barometric sensor. This allows for switching between gauge pressures to absolute pressures.


    Standalone Solution
    Typical pressure controllers will require a nitrogen bottle for the gas supply which make it difficult to move the controller around without having to move or connect to another bottle. The Additel 780 is unique in that with the optional electric pump, you can generate pressures to 1,000 psi (70 bar) without the need of a gas bottle.


    Process Functionality
    The 780 Pressure Controller is considered to handle a very wide range of applications which may normally require a pressure calibrator. Built-in capability, includes current and voltage measurement capability, 24 volt loop power, HART® and Profibus PA communication, switch measurement capability, selectable analog output and much more.

  • The 780 Pressure Controller is considered to handle a very wide range of applications which may normally require a pressure calibrator. Built-in capability, includes current and voltage measurement capability, 24 volt loop power, HART® and Profibus PA communication, switch measurement capability, selectable analog output and much more.

  • SpecificationADT780-1KADT780-3K
    Pressure range-14.5 to 1,000 psi (-0.95 to 70 bar)-14.5 to 3,000 psi (-0.95 to 200 bar)[1]
    Control stability0.003% FS (stability based on % FS of control sensor range)
    Precision (includes 1 year stability)See pressure range table
    MediaClean gas
    Over-range indication103% to 120% (based on sensor)
    Resolution4, 5, 6, or 7 digits (user selectable)[2]
    Pressure typeGauge, Absolute
    Warm up time15 minutes
    Measurement unitsPa, hPa, kPa, MPa, bar, mbar, psi, mmHg@0°C,cmHg@0°C, mHg@0°C, inHg@0°C, inH2O@4°C, mmH2O@4°C, cmH2O@4°C, mH2O@4°C, mmH2O@20°C, cmH2O@20°C, mH2O@20°C, kg/m2, kg/cm2, mtorr, torr, atm, lb/ft2, tsi, user selectable
    Minimum control pressure[3]0.0001 psi0.001 psi
    BAROMETRIC MEASUREMENT SPECIFICATION
    Sensor/ModeGaugeAbsolute
    ADT160-02-CPXXN/A40 Pa[1]
    ADT160-01-APXXQ3 Pa[2]N/A
    Barometer range (60~110)kPa, the accuracy is 40 Pa (11 Pa is optional) [1] 40 Pa uncertainty (k=2) includes calibration uncertainty, linearity, and long term stability (<30 Pa per year). Barometer range of 60 to 120 kPa. [2] Combined linearity, hysteresis, and repeatability. Add 3 Pa when used in gauge mode. When using the ADT160-01-APXXQ sensors with the ADT780 controller in gauge mode, regular zeroing will realign the barometric reading to provide the most accurate result.
    ELECTRICAL MEASURE SPECIFICATIONS
    SpecificationRangeResolutionAccuracy
    Volts DC-30 to 30 V0.1 mV±0.01% rdg + 1.5 mV
    -300 to 300 mV1 μV±0.01% rdg + 15 μV
    Current DC-30 to 30 mA0.1 μA±0.01% rdg + 1.5 μA
    Switch testIf the switch has detected voltage, the range is from 3 – 24 V
    DC 24V output24V ± 0.24V, max 30 mA
    PHYSICAL SPECIFICATIONS
    SpecificationADT780-1K & ADT780-3K
    Power100 to 240V, 50/60 Hz
    Pressure portsG1/8 F
    Storage temperature-20°C to 70°C
    Operating environment0-90% RH non-condensing
    Display7 inch (17.8 cm) color, touch screen display
    Weight33 lbs (15 kg)
    Dimensions (DWH)16.5 x 17.3 x 5.2 inch (419 x 440 x 132 mm)
    MountingStandard desktop, optional rack mount kit
    Shock4G
    Vibration1G 10Hz~500Hz
    OTHER SPECIFICATIONS
    WiFi specifications802.11 b, g, and n
    VentFront panel vent and safety release button (only for 780S)
    CommunicationsRS232,USB,LAN,WIFI
    Stability indicatorUser selectable
    Operating modesControl, measure, and vent
    Display modesController – show pressure indication and control
    LocalizationEnglish, Chinese (simplified), German, Spanish, French, Italian, Portuguese, Russian, Japanese
    ConformityCE
    Calibration certificationAccredited certificate of calibration with data included
    Warranty1 year
    160A for 780 & 780S
    P/NPressure RangeMediaAccuracy[2]Pressure RatingSuggested Controller CompatibilityInternal/ External Use
    psiabar.aBurstOver Pressure
    Precision Sensors[1]
    AP15R0 to 150 to 1G0.01% FS2x1.2xADT780-1K onlyBoth
    AP30R0 to 300 to 2G0.01% rdg or 0.003% FS whichever is greater2x1.2xADT780-1K onlyBoth
    AP50R0 to 500 to 3.5G0.01% rdg or 0.003% FS whichever is greater2x1.2xADT780-1K onlyBoth
    AP100R0 to 1150 to 8G,L0.01% rdg or 0.003% FS whichever is greater2x1.2xADT780-1K onlyBoth
    AP300R0 to 3150 to 21G,L0.01% rdg or 0.003% FS whichever is greater2x1.2xBothBoth
    AP500R0 to 5150 to 36G,L0.01% rdg or 0.003% FS whichever is greater2x1.2xBothBoth
    AP1KR0 to 1,0150 to 71G,L0.01% rdg or 0.003% FS whichever is greater2x1.2xBothBoth
    AP2KQ0 to 2,0150 to 141G0.005% rdg + 0.005% FS3x1.1xADT780-3K onlyBoth
    AP3KQ0 to 3,0150 to 201G0.005% rdg + 0.005% FS3x1.1xADT780-3K onlyBoth
    [1] Contact Additel for other range options. [2] Accuracy includes calibration uncertainty, linearity and long-term stability. **Low pressure sensors (ADT160A-XX-DPX) available for low pressure and differential pressure measurement. Also available for low pressure control (gauge mode only). Ranges from ±1 inH20 (2.5 mbar) to 300 inH2O (700 mbar).
    160A for 780 & 780S
    P/NPressure RangeMediaAccuracyPressure RatingSuggested Controller CompatibilityInternal/ External Use
    psigbar.gBurstOver Pressure
    Standard Pressure Sensors
    CP10±10±0.7G0.02% FS3x1.2xADT780-1K onlyBoth
    CP15±15±1.0G0.02% FS3x1.2xADT780-1K onlyBoth
    CP30-15 to 30-1 to 2.0G0.02% FS3x1.2xADT780-1K onlyBoth
    CP50-15 to 50-1 to 3.5G0.02% FS3x1.2xADT780-1K onlyBoth
    CP100-15 to 100-1 to 7.0G,L0.02% FS3x1.2xADT780-1K onlyBoth
    CP300-15 to 300-1 to 20G,L0.02% FS3x1.2xBothBoth
    CP500-15 to 500-1 to 35G,L0.02% FS3x1.2xBothBoth
    CP600-15 to 600-1 to 40G,L0.02% FS3x1.2xBothBoth
    CP1K-15 to 1,000-1 to 70G,L0.02% FS3x1.2xBothBoth
    CP2K-15 to 2,000-1 to 140G,L0.02% FS3x1.2xADT780-3K onlyBoth
    CP3K-15 to 3,000-1 to 200G,L0.02% FS3x1.2xADT780-3K onlyBoth
    CP5K-15 to 5,000-1 to 350G,L0.02% FS3x1.2xN/AExternal
    CP10K-15 to 10,000-1 to 700G,L0.02% FS2x1.2xN/AExternal
    160A GENERAL SPECIFICATIONS
    Standard AccuracyPrecision Accuracy
    CPXXXAPXXQAPXXXR
    Operating temperature-10°C to 50°C (14°F to 122°F)15°C to 25°C (59°F to 77°F)10°C to 30°C (50°F to 86°F)
    Storage temperature-20°C to 70°C (-4°F to 158°F)-20°C to 70°C (-4°F to 158°F)
    Relative humidity95% RH90% RH
    Pressure connections (for external use only)1/4NPT, 1/2NPT, 1/4BSP, 1/2BPS, M20x15
    Enclosure (for external use only)SS enclosure
    Dimensions (Dia x H)33 mm x 123 mm (1.3” x 4.84”)44 mm x 220mm (1.7” x 8.7”)44 mm x 195 mm (1.7” x7.7”)
    Weight0.4 kg (0.99 lb)0.8 kg (1.8 lb)0.7 kg (1.5 lb)
    ADT780-1K-EPUMP
    Pressure range-13.5 to 1,080 psi (-0.95 to 75 bar)
    Weight69.6 lbs (31.6 kg)
    Size(WHD)17.3 X 9.3 X 21.7 inch (440 X 235X 550 mm)
    Power100 to 240V
    MediaAir
    Volume150 mL
    Outlet portG1/8F
    Storage temperature-20°C to 70°C
    Operating environment0-90% RH non-condensing
    MountingStandard desktop, optional rack mount
    Control operation2 LED displays with pressure limit settings
    Typical max pressure timeApprox 2 min with ADT780 Controller

  • Documents:

    File nameUpload DateSize
    Additel 780 series pressure controller data sheet 2018.11.13 1.2 MB
    Additel 780 pressure controller User Manual 2018.3.27 6.6 MB
    780-1K-EPUMP Electrical Pump User Manual 2018.3.27 1.3 MB

    Update Packages:

    File name Upload Date Size
    Additel 780 Firmware Update Package 4.4 MB