At Additel, innovation and continuous improvement are part of our company's culture and the products we introduce. When we set out to deliver the Additel 761A series calibrators, we knew we needed to provide breakthrough improvements and additional value to the existing line of calibrators (Additel 761 series). The ADT761A has many improvements: increased pressure range to 1,000 psi (70 bar), removable internal pressure modules, optional precision models to 0.01%FS, increased speed to pressure, ability to read two external pressure modules, touch screen display, Wi-Fi, Bluetooth, and Ethernet communications, double the original battery life, and more!
ADT761A-BP
The Additel 761A-BP is designed for calibration of barometer sensors. With a range of 100 to 1200 hPa and an accuracy of 0.01%FS, this unit is ideal for calibration on the bench or in the field. -
-
Pressure Specifications Specification 761A-LLP 761A-D 761A-500 761A-1K 761A-APXR 761A-BP Max Pressure Range 30 inH2O (75 mbar) 35 psi (2.5 bar) 500 psig (35 bar.g) 1,000 pisg (70 bar.g) Dependent on APXR 1,200 hPasensor Min Pressure Range -30 inH2O (-75 mbar) -13.5 psi (-0.95 bar) -13 psi (-0.9 bar) -13 psi (-0.9 bar) Dependent on APXR 100 hPasensor Accuracy(1) 0.05%FS 0.02%FS 0.02%FS 0.02%FS 0.01%FS 0.01%FS Stability 0.005%FS or 0.05 Pa 0.005%FS 0.005%FS 0.005%FS 0.003%FS 0.02 hPa Pressure Type Differential, Gauge Differential, Gauge Gauge, Absolute Gauge, Absolute Gauge, Absolute Absolute Over Range Indication 1.2 Resolution 6 digits Measurement Units Pa, hPa, kPa, mPa, bar, mbar, psi, mmHg@0°C, cmHg@0°C, mHg@0°C, inHg@0°C, inH2O@4°C, mmH2O@4°C, cmH2O@4°C, mH2O@4°C, mmH2O@20°C, cmH2O@20°C, mH2O@20°C, inH2O@20°C, inH2O@68°F, kg/m2, mtorr, torr, lb/ft2, tsi, custom Barometric Accuracy N/A N/A 60 Pa 60 Pa 60 Pa N/A [1] One year accuracy (including 1 year stability). FS specification applies to the span of the module range. [2] Specification based on gauge measurement. An additional 60 pa uncertainty will need to be included when measuring in absolute mode. Applicable only for use with the ADT761A-500 and ADT761A-1K * Additel 761A calibrators support 160A intelligent digital pressure modules that are available for gauge, vacuum and absolute pressure from -15 psi to 60,000 psi (-1 bar to 4200 bar).For detailed specs, see the 160A pressure module data sheet. Electrical Specifications Specification Range Resolution Accuracy Note mA Measure Auto-ranging, ±25 or ±50 mA 0.1 μA 0.008%RD+0.004%FS Impedance<10Ω V Measure ±300 mV 1 μV 0.008%RD+0.002%FS Impedance >1GΩ Auto-ranging, ±5, ±12 or ±30 V ±5V:20 μV 0.008%RD+0.002%FS Impedance >1MΩ ±12V:100 μV ±30V:100 μV Loop Power Source 24 V N/A ±0.24 V 50 mA (Max Loading) mA Source 0 to 2.5 mA or 2.5 to 25 mA 0-2.5 mA: 0.05μA 0-2.5 mA: 0.008%RD+0.1 μA 20 mA @ 1KΩ 0-25 mA:0.5μA 0-25 mA: 0.008%RD+1.0 μA Power Source 16 to 30 V 1 V ±0.5 V(24 V: ±0.24 V) 70 mA (Max Loading) V Source 0 to 16 V 250 μV 0.008%RD+0.002%FS Pressure Switch Mechanical Switch, Live Mechanical Switch, NPN Switch, PNP Switch N/A N/A Response time<10 ms. If the switch is live, voltage range will be (3-30) V Temperature Compensation 5°C to 35°C Temperature Coefficient Outside of 5°C to 35°C: <±0.0005%RD+0.0001%FS/°C Misuse Protection Up to 30 V on any two sockets Internal Module Specification and Compatibility Model inH2O mbar Media Accuracy (%FS)(1) Burst Pressure 761A-LLP 761A-D 761A-500 761A-1K ADT155-20-DP025 ±0.25 ±0.62 G 0.2 (2) 100x • ADT155-10-DP050 ±0.5 ±1.25 G 0.1(3) 100x • ADT155-05-DP1 ±1 ±2.5 G 0.05(4) 100x • ADT155-05-DP2 ±2 ±5 G 0.05(4) 100x • ADT155-05-DP5 ±5 ±10 G 0.05(4) 50x • ADT155-05-DP10 ±10 ±25 G 0.05(4) 20x • • ADT155-05-DP20 ±20 ±50 G 0.05 20x • • ADT155-05-DP30 ±30 ±75 G 0.05 20x • • ADT155-05-DP50 ±50 ±125 G 0.05 3x • ADT155-02-DP100 ±100 ±250 G 0.02 3x • ADT155-02-DP150 ±150 ±350 G 0.02 3x • ADT155-02-DP300 ±300 ±700 G 0.02 3x • ADT155-02-DP400 -380 to 400 -950 to 1K G 0.02 3x • ADT155-02-DP800 -380 to 800 -950 to 2K G 0.02 3x • ADT155-02-DP1K -380 to 1K -950 to 2.5K G 0.02 3x • [1] FS specification applies to the span of the module range. Accuracy includes one-year stability, except for DP025 to DP10 modules. [2] Accuracy is a 6 months spec, 1-year long-term drift is 0.2%FS. [3] Accuracy is a 6 months spec, 1-year long-term drift is 0.1%FS. [4] Accuracy is a 6 months spec, 1-year long-term drift is 0.05%FS. [5] Specification based on gauge measurement. An additional 60 pa uncertainty will need to be included when measuring in absolute mode.Applicable only for use with the ADT761A-500 and ADT761A-1K. General Specifications Specification Description User Interface Color touch screen and/or keypad operation Four total: one electrical, high or low internal Channels pressure module, two external pressure modules Enclosure IP Rating IP31 Battery Rechargeable Li-Ion battery, typically 16 hours of operation, recharges in less than 5 hours. Power Rechargeable Li-Ion battery, external power 110/220 V power adapter 27 V Display 7” TFT touch screen 800 x 480 color Communications USB, LAN, Bluetooth, Wi-Fi Weight <17.52 lb (7.95 KG) Size 11.77 x 7.60 x 7.56 in (299 x 193 x 192 mm) Certification ISO17025 accredited certificate of calibration with NIST-traceable data Data Storage 6 GB Data Logging Up to 1,000,000 readings (data and time stamped) HART Communicator Read, configure and calibrate HART devices - DD files updated periodically Task documentation Up to 1000 tasks can be stored with data Automation Functions Switch test, auto step, leak test User Interface Localization English, German, French, Italian, Spanish, Portuguese, Chinese, Japanese, and Russian Pump life >100,000 cycles Environmental Specifications Operating temperature: 32 °F to 122 °F( 0°C to 50°C) Compensated temperature: 32 °F to 122 °F( 0°C to 50°C) Storage temperature: -20°C to 60°C ( -4 °F to 120 °F) Humidity: <90%, non-condensing Vibration and Shock Vibration: 4g (20 to 2,000 Hz) Shock: 8g, 1 meter drop test Compliance CE Software Compatibility ACal, Additel Land and Additel Link for access via mobile application Warranty 1 year Pressure Range Type Model Pressure Range High-Range Pressure Module Low-Range Pressure Module Range Accuracy Range Accuracy Low/Differential Pressure ADT761A-LLP -75 to 75 mbar (-30 to 30 inH2O) DP30: -75 to 75 mbar (-30 to 30 inH2O) 0.05%FS User selectable from DP20 to DP025 See Internal Module Table ADT761A-D -0.95 to 2.5 bar (-13.5 to 35 psi) CP35: -0.95 to 2.5 bar (-13.5 to 35 psi) 0.02%FS User selectable from DP10 to CP30 See Internal Module Table Standard Accuracy Gauge/Absolute Pressure ADT761A-500 -0.9 to 35 bar (-13 to 500 psi) CP500: -0.9 to 35 bar (-13 to 500 psi) 0.02%FS User selectable from CP10 to CP300 See Internal Module Table ADT761A-1K -0.9 to 70 bar (-13 to 1000 psi) CP1K: -0.9 to 70 bar (-13 to 1000 psi) 0.02%FS User selectable from CP30 to CP600 See Internal Module Table Precision Accuracy Gauge/Absolute Pressure ADT761A-30R -0.95 to 1 bar (-13.5 to 15 psi) AP30R: -0.95 to 1 bar (-13.5 to 15 psi) 0.01%FS Single sensor only ADT761A-50R -0.95 to 2.5 bar (-13.5 to 35 psi) AP50R: -0.95 to 2.5 bar (-13.5 to 35 psi) 0.01%FS Single sensor only ADT761A-100R -0.95 to 7 bar (-13.5 to 100 psi) AP100R: -0.95 to 7 bar (-13.5 to 100 psi) 0.01%FS Single sensor only ADT761A-300R -0.9 to 21 bar (-13 to 300 psi) AP300R: -0.9 to 21 bar (-13 to 300 psi) 0.01%FS Single sensor only ADT761A-500R -0.9 to 35 bar (-13 to 500 psi) AP500R: -0.9 to 35 bar (-13 to 500 psi) 0.01%FS Single sensor only ADT761A-1KR -0.9 to 70 bar (-13 to 1000 psi) AP1KR: -0.9 to 70 bar (-13 to 1000 psi) 0.01%FS Single sensor only -
Documents:
File name Upload Date Size Additel 761A Automated Pressure Calibrator Datasheet 2019.8.26 0.8 MB ADT761A Selection Guide.pdf 2019.8.26 0.1 MB Additel 761A User Manual 2018.3.28 2.2 MB Additel 761A Battery MSDS 2020.6.11 5.2 MB Additel Hart Communication Manual 2020.7.28 0.4 MB Update Packages:
File name Upload Date Size Additel 761A Firmware Package 2020.2.18 11.4 MB Hart Pressure Product DD Library 2020.8.20 139.4 MB -
Selection Guide Models 761A-LLP 761A-D 761A-500 761A-1K 761A-APXR 761A-BP Pressure Range -75 to 75 mbar (-30 to 30 inH2O) -0.95 to 2.5 bar.g (-13.5 to 35 psig) -0.90 to 35 bar.g (-13 to 500 psig) -0.90 to 70 bar.g (-13 to 1000 psig) -0.90 to 70 bar.g (-13 to 1000 psig) 100 to 1,200 hPa Control Stability <0.005%FS or 0.05Pa <0.005%FS <0.005%FS <0.005%FS <0.003%FS <0.02 hPa Number of Internal Modules 2 2 2 2 1 1 Removable Internal Modules • • • • Differential Pressure • • Gauge Pressure • • • • • Absolute Pressure • • • • Barometric Pressure • • • • Two External Pressure Modules • • • • • • Built-in Electric Pump • • • • • • Source/Simulate mA • • • • • • Measure mA or V • • • • • • 24V Loop Power • • • • • • Pressure Switch Test • • • • • • HART/Profibus Communication • • • • • • Task Documentation • • • • • • Data Logging • • • • • •