Explore the Interface

logo Additel

รุ่น   681


Description Additel 681

Digital Pressure Gauges
  • Pressure ranges to 60,000 psi (4,200bar)
  • 0.02%, 0.05%, 0.1% or 0.2% FS or 0.1% RD accuracy
  • % pressure indication with fan-shaped graph scale for visual reference
  • Fully temperature compensated accuracy
  • Panel mount gauges are available (optional)
  • Intrinsically safe version 681IS (optional)
  • Data logging (optional)
  • IP67 rated (681IS)
  • ISO 17025-accredited calibration w/data included


  • With advanced microprocessor technology and state-of-the-art silicon pressure sensors, the 681 series digital pressure gauges provide an accurate, reliable, and economic solution for a wide range of pressure applications. They are loaded with functionality and remarkably easy to use. To reach the best performance, every silicon pressure sensor in our gauges is specially aged, tested and screened before assembly. At Additel, fully temperature compensated accuracy means every sensor is pressure tested at several environmental temperatures from -10°C to 50°C. With this test data individual coefficients are generated and stored in the gauge characterizing its performance over the full temperature compensated range. And now the ADT681IS comes with an IP67 rating meaning it is dust resistant and water proof, submerible in 1 meter of water.
    Designed to fit your need
    Additel pressure gauges give you the widest variety of sensor choices on the market. Whether you require low inches of water measurement or very high pressure measurement, we have a gauge that will meet your need. We offer sensors which are ±1 inH2O (±2.5 mbar) to 60K psi (4,200 bar) and everything in between.
    Do your applications require you to measure both positive and vacuum pressure? Our compound gauges do not compromise accuracy and provide you with the same high accuracy specification on both positive and vacuum pressures. We offer a wide variety up to 300 psi (20 bar). If you need a higher range, just contact us and we can likely customize one to meet your need. We also offer absolute pressure sensors to 5K psi (350 bar) and a full range of differential pressure sensors from ±1 inH2O (±2.5 mbar) to ±300 inH2O (±700 mbar). Are you looking for a pressure gauge to use in hazardous areas? Our certified (ATEX, CSA US & IECEx) intrinsically safe models (681IS) are designed for pressure measurement in hazardous areas.
    If you need to panel mount our sensors, we offer the option (see ordering information) for a back-mounted pressure port and gauge housing designed to fit in a panel. And most recently, we've added the option to do stand-alone data logging with the 681. Now you can record more than 21,000 records internal to the 681 series. Each record includes date, time, pressure and temperature readings. Download the logged data with our free Additel/Land software or you can purchase our Additel/Log II for real-time logging and data analysis. The 681 series digital pressure gauges are unmatched in performance and reliability. Best of all, they are very affordable.

    • Precision pressure measurement
    • Field gauge calibration
    • Pressure safety valve (PSV) testing
    • Pressure regulator testing

  • Gauge Pressure[1]
    P/N Pressure Range Media[2] Accuracy Burst Pressure
    psi bar %FS %RD
    V15 -15 -1 G 0.02 (0.05, 0.1, 0.2) N/A
    GP2 2 0.16 G 0.05 (0.1, 0.2) N/A
    GP5 5 0.35 G,L 0.05 (0.1, 0.2) 0.1
    GP10 10 0.7 G,L [3] 0.02 (0.05, 0.1, 0.2) 0.1
    GP15 15 1 G, L[3] 0.02 (0.05, 0.1, 0.2) 0.1
    GP30 30 2 G, L[3] 0.02 (0.05, 0.1, 0.2) 0.1
    GP50 50 3.5 G, L 0.02 (0.05, 0.1, 0.2) 0.1
    GP100 100 7 G, L 0.02 (0.05, 0.1, 0.2) 0.1
    GP150 150 10 G, L 0.02 (0.05, 0.1, 0.2) 0.1
    GP300 300 20 G, L 0.02 (0.05, 0.1, 0.2) 0.1
    GP500 500 35 G, L 0.02 (0.05, 0.1, 0.2) 0.1
    GP600 600 40 G, L 0.02 (0.05, 0.1, 0.2) 0.1
    GP1K 1000 70 G, L 0.02 (0.05, 0.1, 0.2) 0.1
    GP2K 2000 140 G, L 0.02 (0.05, 0.1, 0.2) 0.1
    GP3K 3000 200 G, L 0.02 (0.05, 0.1, 0.2) 0.1
    GP5K 5000 350 G, L 0.02 (0.05, 0.1, 0.2) 0.1
    GP10K 10000 700 G, L 0.02 (0.05, 0.1, 0.2) 0.1
    GP15K 15000 1000 G, L 0.05 (0.1,0.2) 0.1
    GP20K 20000 1400 G, L 0.05 (0.1,0.2) N/A 1.5×
    GP25K 25000 1600 G, L 0.1 (0.2) N/A 1.5×
    GP30K 30000 2000 G, L 0.1 (0.2) N/A 1.5×
    GP36K 36000 2500 G, L 0.1 (0.2) N/A 1.5×
    GP40K 40000 2800 G, L 0.1 (0.2) N/A 1.35×
    GP50K 50000 3500 G, L 0.1 (0.2) N/A 1.2×
    GP60K 60000 4200 G, L 0.1 (0.2) N/A 1.1×
    [1] Sealed gauge pressure for above 1,000 psi [2] G=Gas, L=Liquid [3] 0.02% FS for gas media only
    Barometric Pressure
    P/N Pressure Range Media Accuracy Burst Pressure
    Low High
    BP 60 kPa 110 kPa G 40 Pa
    Absolute Pressure
    P/N Pressure Range Media[1] Accuracy(%FS) Burst Pressure
    psi bar
    AP5 5 0.35 G 0.1 (0.2)
    AP10 10 0.7 G 0.1 (0.2)
    AP15 15 1 G 0.1 (0.2)
    AP30 30 2 G 0.1 (0.2)
    AP50 50 3.5 G 0.1 (0.2)
    AP100 100 7 G, L 0.05 (0.1, 0.2)
    AP300 300 20 G, L 0.05 (0.1, 0.2)
    AP500 500 35 G, L 0.05 (0.1, 0.2)
    AP1K 1000 70 G, L 0.05 (0.1, 0.2)
    AP3K 3000 200 G, L 0.05 (0.1, 0.2)
    AP5K 5000 350 G, L 0.05 (0.1, 0.2)
    [1] G=Gas, L=Liquid
    Differential Pressure
    P/N Pressure Range Media Accuracy (%FS)[1] Burst Pressure Static Pressure Range
    (inH2O) (mbar)
    DP1 ±1 ±2.5 G 0.05[2] 100× ±10 psi
    DP2 ±2 ±5.0 G 0.05[2] 100× ±10 psi
    DP5 ±5 ±10 G 0.05[2] 50× ±10 psi
    DP10 ±10 ±25 G 0.05[2] 20× ±10 psi
    DP20 ±20 ±50 G 0.05 20× ±10 psi
    DP30 ±30 ±75 G 0.05 20× ±10 psi
    DP50 ±50 ±160 G 0.05 ±10 psi
    DP100 ±100 ±250 G 0.02 ±15 psi
    DP150 ±150 ±350 G 0.02 (0.05) 50 psi
    DP300 ±300 ±700 G 0.02 (0.05) 50 psi
    [1] FS specification applies to the span of the range. Accuracy includes one year stability. [2] 0.05%FS accuracy (incl 6 months stability). One year accuracy is 0.05%FS calibration accuracy combined with 0.05%FS one year stability.
    Compound Pressure
    P/N Pressure Range Media[1] Accuracy (%FS)[1] Burst Pressure
    psi bar %FS[2] %RD
    CP2 ±2 ±0.16 G 0.05 (0.1,0.2) N/A
    CP5 ±5 ±0.35 G 0.02 (0.05,0.1,0.2) 0.1
    CP10 ±10 ±0.7 G 0.02 (0.05,0.1,0.2) 0.1
    CP15 ±15 ±1 G 0.02 (0.05,0.1,0.2) 0.1
    CP30 -15 to 30 -1 to 2 G 0.02 (0.05,0.1,0.2) 0.1
    CP100 ﹣15 to 100 ﹣1 to 7 G, L 0.02 (0.05,0.1,0.2) 0.1
    CP300 ﹣15 to 300 ﹣1 to 20 G, L 0.02 (0.05,0.1,0.2) 0.1
    [1] G=Gas, L=Liquid [2] FS specification applies to the span of the range.
    General Specifications
    Model ADT681 ADT681IS
    Description Digital Pressure Gauge Intrinsically Safe Digital Pressure Gauge
    Intrinsic Safety & European Compliance CE marked
    Ɛx II 1G EX ia IIC T4 Ga
    ATEX certified intrinsically safe
    Accuracy (For detailed accuracy, please see pressure range table) 681(IS)-02 : 0.02% of full span
    681(IS)-05 : 0.05% of full span
    681(IS)-10 : 0.1% of full span
    681(IS)-20 : 0.2% of full span
    681(IS)-RD : 0.02% of full span (0% to 20% of Range), 0.1% of reading (20 to 100% of Range), 0.25% of full span (Vaccuum)
    Gauge Types Gauge pressure, Compound pressure, Absolute pressure, Differential pressure, Barometric pressure
    Fan-shaped Graph Scale Similar to analog dials, including pressure swing, % indication with fan-shaped graph scale for visual reference, low/high alarm
    Display Description: 5 full digit FSTn LCD
    Display rate: 3 readings per second (Default setting). Adjustable from 10 readings per second to 1 reading every ten seconds
    Numeral display height: 16.5mm (0.65'')
    Pressure Units Pa, kPa, MPa, psi, bar, mbar, kgf/cm2, inH2O @4°C, mmH2O @ 4°C, inHg@0°C, mmHg @0°C
    Enviromental Compensated Temperature: 14°F to 122°F (-10°C to 50°C)
    Operating Temperature: 14°F to 122°F (-10°C to 50°C)
    Storage Temperature: -4°F to 158°F (-20°C to 70°C)
    Humidity: <95%
    Pressure Port 15,000 psi: 1/4NPT male, 1/2NPT male, 1/4BSP male, 1/2BSP male, M20x1.5 male
    >15,000 psi: 1/4HP female (Autoclave F-250-C, 9/15" - 18 UNF-2B) or 1/4HP male (Autoclave M-250-C, 9/15" - 18 UNF-2A)
    Differential Pressure: 0.236 inch (Ø6 mm) test hose
    Other connectsion available upon request
    Power Battery: One 9V alkaline battery (included)
    Battery Life (High Power Mode): 320 hours
    Battery Life (Low Power Mode): 300 hours (10 readings/s), 500 hours (3 readings/s), or 4000 hours (1 reading / 10s)
    Power auto-off: 60 minutes power auto-off. Auto-off may be disabled.
    External power: 100/220V external power adapter (optional)
    Enclosure Case material: Aluminum Alloy
    Wetted parts: 316L SS
    Dimension: Ø110mm X 35mm depth X 176mm height (panel mount gauge: Ø140mm X 86mm depth)
    Weight: 0.6kg
    Compliance Protection Level: IP67 (available for 681IS GP15-60K)
    Vibration: 5g (20-2000 Hz)
    Shock Resistance: 100g/11ms
    Data Logging (Available with data logging option ADT681-...-DL) Data storage: 21,843 records (each record includes data, time, pressure, and temperature)
    Rate: user -selectable from 1 to 99,999 second intervals
    Communication RS232 (Do not use the RS-232 connector in a hazardous atmosphere)
    Warranty 1 year