Additel 160A 01

logo Additel

รุ่น  209/210

Description Additel 209/210

Series Loop Calibrator
  • Accuracy to 0.01% of reading
  • Measure, source, or simulate loop current
  • Measure DC volts
  • Simultaneously mA and % span display
  • Switch functionality
  • Selectable ramp and step functions
  • HART 250Ω resistor in series with 24V loop
  • ISO 17025-accredited calibration w/data included

  • The new Additel 209 and 210 loop calibrator series combine ease of use and functionality, making them the ideal tools to troubleshoot your process loop. The ADT209 has an accuracy of 0.03% of reading whereas the ADT210 holds an accuracy of 0.01% of reading. If you want to source, simulate or simply measure the Additel Loop calibrator series will fit your need. The ADT209 and ADT210 allow for measurement of current, voltage and a switch. You can also simulate or source mA or a process transmitter. With a push of a button, you can switch to zero and span values, auto ramp, and auto step throughout the range. Each loop calibrator has a large, easy to read screen which simultaneously displays the measurement with the % of span.
  • The ADT209 and ADT210 allow for measurement of current, voltage and a switch. You can also simulate or source mA or a process transmitter. With a push of a button, you can switch to zero and span values, auto ramp, and auto step throughout the range.
  • Electrical Measure Specifications
    RangeResolutionADT209 AccuracyADT210 Accuracy
    Voltage DC10 to 30 V1 mV0.03%RD + 2mV0.01%RD + 2mV
    Current DC20 to 24 mA1 µA0.03%RD + 2µA0.01%RD + 2µA
    Switch testInput resistance more than 500 MΩ Trigger level: low level <0.3V; high level: >2V
    [1] 1MΩ input resistance [2] Loop transmitter current measure: 700Ω maximum
    Electrical Source Specifications
    RangeResolutionADT209 AccuracyADT210 Accuracy
    Current DC0 to 24 mA1 µA0.03%RD + 2µA0.01%RD + 2µA
    Source mode700Ω/20 mA maximum
    Sink modeExternal loop voltage nominal 24V, maximum 30V, minimum 12V
    General Specifications
    Voltage limit30V between terminals or between terminals and ground
    Measurement FunctionsAuto step, auto ramp, span step
    DisplayVA LCD display. 2.04 x 2.04 in (52 x 52 mm)
    Loop power24V
    Over-voltage protection30V DC (240V AC)
    Overload current protection33 mA DC
    Storage temperature-20°C to 70°C
    Working Environment-10 to 50°C, 95%RH
    Working Altitude<3,000 m
    Vibration/shockRandom 2G 5 to 500Hz 1 meter drop
    PowerOne 9V alkaline battery (ANSI/NEDA 1604A or IEC) DC9V optional adapter available
    Battery life (typical)Output mode: 18 hours (12 mA/500Ω) Measure mode: 50 hours
    Size (LxWxH)163 x 83 x 41 mm
    Weight350 g
    Calibration CertificationISO17025 Accredited certificate of calibration with data
    Compliance CertificationCE
    Warranty3 years
  • Documents:

    File nameUpload DateSize
    Additel 209 and 210 Series Loop Calibrator Data Sheet 2018.3.28 0.2 MB
    Additel 209 and 210 Series Loop Calibrator User Manual 2018.3.28 0.6 MB
    Additel Multifunction Process Calibrator Selection Guide 2019.5.21 0.2 MB

    Update Packages:

    File name Upload Date Size
    0 MB
  • Ordering Information

    Model Numbers
    • ADT209
    • ADT210

    Standard Accessories
    Test lead set: 1 set
    Alligator clips: 2 pcs
    User manual:1 pc
    9V Alkaline battery:1 pc

    Optional Accessories
    Model 9812: 110V/220V external power adapter (DC 9V)